EasyManua.ls Logo

Covidien WarmTouch - B.2 Error Codes and Suggested Resolutions

Covidien WarmTouch
314 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Error Codes
B-2 Service Manual
B.2 Error Codes and Suggested Resolutions
Error Categories:
A/D — Errors associated with or reported by the A/D microprocessor.
POST — Errors occurring during power-on self-test.
UI — Errors associated with or reported by the UI microprocessor.
Note:
For most errors, the initial corrective action is to power-cycle the warming unit. If the
error persists, continue with the recommendations listed for the error.
003
Category: A/D
Alarm Priority: Low
Description: Checksum error on code in flash. A/D flash is defective.
Resolution: Try reloading software to the unit. See page 7-7.
If this does not resolve the issue, replace the power PCBA. See page 10-31.
004
Category: A/D
Alarm Priority: Low
Description: One of the following conditions occurred:
Temperature was more than ±1°C from the setpoint for at least 30 consecu-
tive seconds after the setpoint was reached.
Temperature did not reach the specified setpoint within 10 minutes.
Resolution: Check for:
Blockages in the air intake, hose, and filter.
This error may also appear due to issues at the customer site:
Low mains voltage
Ambient temperature outside of specification.

Table of Contents