ESD (Electrostatic Discharge) Precautions
Service Manual 1-7
1.3 ESD (Electrostatic Discharge) Precautions
Caution:
Observe ESD (electrostatic discharge) precautions when servicing the
warming unit.
To avoid damaging ESD-sensitive components while service the warming unit,
follow all appropriate ESD guidelines. These guidelines include:
Using an ESD wrist strap, properly connected to a reliable ground.
Using an anti-static mat.
Handling ESD-sensitive components properly. (Do not touch connection points,
connector pins, leads, or terminals.)
Keeping non-conducting materials (plastic containers, foam cups, synthetic cloth-
ing, cellophane tape, etc.) away from the work area.
1.4 Tools Required for Service
Electrical leakage tester
Ground bond tester
Ohmmeter or digital multi-meter with resistance function
Temperature probe (see page 7-2 for requirements)
Torque driver with:
#1 Phillips bit
#2 Phillips bit
11/32” hex socket
10mm hex socket
(See Table 1-2 on page 1-8 for torque requirements.)
10 mm open-end or box wrench
70% isopropyl alcohol solution and soft cloth for cleaning
Small flat-blade screwdriver
Pliers