Operation
4-8 Service Manual
4.6.4 Returning to Standby Mode
Use Standby mode when you are finished using the warming system but do
not wish to unplug it.
To return to Standby mode during operation, press the On/Standby key. The
heater and fan shut off, the display goes blank, and the status LED turns
amber.
4.6.5 If Power is Interrupted...
If power is interrupted during operation, the warming unit responds as fol-
lows, based on the duration of the power loss:
• Power restored within approximately 15 seconds — The warming unit
powers up, performs a power-on self-test (POST), and resumes operation in the
same mode as before the power loss.
• Power restored after approximately 15 seconds — The warming unit
powers up in Standby mode.
If the warming unit powers back up in Standby mode, press the On/Standby
key to enter Ready mode, then select a temperature setting to resume opera-
tion.
4.7 Powering Off the Warming Unit
To power off the warming unit:
1. Press the On/Standby key. The fan and heater (if running at the time) turn off.
The display goes blank, and the status LED turns amber.
2. To completely remove power from the warming unit, unplug the unit’s power
cord from the outlet.