Performance Verification
7-2 Service Manual
7.3 Required Test Equipment
Temperature probe (either handheld meter or USB-based):
Accuracy of probe must be ±0.25°C or better over the range 10°C to 60°C.
Probe cable must be long enough to reach from computer to hose duct
adapter screen inside compressed hose.
Electrical leakage tester
Ground bond tester
7.4 Service Screen Access
The warming unit provides screens for performing functions restricted to
service personnel. These screens are protected by a key code.
When you see the message “Enter Key Code” press the keys indicated in
Figure 7-1 simultaneously to access these screens.
Figure7-1.Key Code Screen