Installing the Warming Unit
Service Manual 3-5
3.4 Installing the Warming Unit
The warming unit can be used as a standalone device, or it can be mounted
on an IV pole, bed, or optional transport cart. The following sections provide
mounting instructions.
3.4.1 IV Pole Installation
The warming unit can be mounted on an IV pole with a maximum diameter of
3.2 cm (1.25 inches) and a minimum diameter of 1.9 cm (0.75 inches).
To avoid tipping, the warming unit must be mounted on the pole with the
handle no higher than 76 cm (30 inches) above the floor.
Requirements:
• Measuring tape
To install the warming unit on an IV pole:
1. On the IV pole, mark the maximum mounting height. The warming unit’s handle
must be no higher than 76 cm (30 inches) above the floor.
2. Ensure that the power cord is secured at the back of the warming unit. Reference
Figure 3-4 on page 3-6.
3. Locate the mounting channel with two clamps at the back of the warming unit.
Do not use the mounting channel with one clamp. Reference Figure 3-4 on
page 3-6.
4. Loosen the knobs on the two clamps and rotate the clamp feet away from the
mounting channel.