User Interface (UI) PCBA Replacement
Service Manual 10-13
WARNING:
Disconnect the warming unit's power cord from the AC power source before
performing this procedure.
To replace the UI PCBA:
1. Follow the procedure Separating the Front and Rear Enclosures on page 9-2.
2. Disconnect the cables from the UI PCBA as follows (Figure 10-11):
a. Slide the locking bar up from the J204 connector, and disconnect the keypad
cable.
b. Open the locking ears at the P204 connector, and disconnect the ribbon
cable.
c. Lift up the locking tab at the J207 connector, and disconnect the display cable.
Note:
Do not remove the USB cable yet (item 2 in Figure 10-11).