Internal Component Replacement (Inside Front Enclosure)
10-50 Service Manual
4. Attach the new ground cable to the equipotential stud as follows:
a. Place the ground cable terminal on the stud.
b. Place the lock washer on top of the terminal.
c. Place the nut on top of the lock washer. Tighten the nut to 3.2 to 3.8 N-m
(28.3 to 33.6 lb-in).
5. Connect the ground cable to the GND terminal on the power supply (Figure 10-
35 on page 10-48). Make sure the connector is fully seated on the terminal, and
that the heat shrink on the adjacent capacitor does not get caught under the con-
nector.
6. Follow the procedure Rejoining the Front and Rear Enclosures on page 9-5.
7. Perform the following tests:
a. Power-On Test on page 7-3.
b. Electrical Safety Tests on page 7-14.