Safety Information
Service Manual 1-5
WARNING:
Using the warming system on patients with transdermal medication patches
may increase the rate of drug delivery, potentially causing harm to the
patient.
WARNING:
Thermally conductive materials, such as water, gel, and similar substances,
can decrease the patient's body temperature when the warming unit is
switched off.
WARNING:
Do not use the warming system during magnetic resonance imaging (MRI)
scanning.
WARNING:
Do not operate the warming system in a stacked configuration with other
equipment.
WARNING:
Do not operate the warming system adjacent to other equipment. If such a
configuration cannot be avoided, first test the warming system in the
intended configuration to verify normal operation.
WARNING:
Use WarmTouch™ blankets only as directed. Carefully follow the Instructions
for Use provided with the blankets regarding proper handling and
positioning.
WARNING:
WarmTouch™ blankets are for single patient use only.
WARNING:
Clean the warming unit after each use, as described in this manual.
WARNING:
If a fault or sudden change in performance occurs in the warming system,
discontinue use. Notify your sales/service center. The unit must be serviced by
qualified personnel using procedures provided in the Service Manual.