Error Codes
B-16 Service Manual
213
Category: UI
Alarm Priority: Low
Description: Read error from RTC.
Resolution: This error may appear due to electromagnetic interference from other devices at
the customer site. Advise the customer to try increasing the distance between
devices, repositioning the cabling, or plugging the devices into separate outlet
circuit branches. See Electromagnetic Compatibility (EMC) on page 13-4 for addi-
tional guidance.
214
Category: UI
Alarm Priority: Low
Description: Write error to RTC.
Resolution: This error may appear due to electromagnetic interference from other devices at
the customer site. Advise the customer to try increasing the distance between
devices, repositioning the cabling, or plugging the devices into separate outlet
circuit branches. See Electromagnetic Compatibility (EMC) on page 13-4 for addi-
tional guidance.
215
Category: UI
Alarm Priority: Low
Description: Unspecified RTC error.
Resolution: This error may appear due to electromagnetic interference from other devices at
the customer site. Advise the customer to try increasing the distance between
devices, repositioning the cabling, or plugging the devices into separate outlet
circuit branches. See Electromagnetic Compatibility (EMC) on page 13-4 for addi-
tional guidance.