Alarms
Service Manual 6-13
To troubleshoot an alarm:
1. If the alarm is currently indicated on the unit’s display, try clearing the alarm as
follows:
a. Power off the warming unit by pressing the key indicated on the alarm screen.
b. Press the On/Standby key to power the warming unit back on.
Note: While a low-priority alarm may be cleared this way, a medium-priority alarm
typically will not be cleared.
2. If possible, check the error codes on the System Information screen (see Serial
Number, Software Version, and Error Codes on page 1-9). Perform the repair(s)
listed for the error code(s) (see Error Codes and Suggested Resolutions on
page B-2).