Performance Tests
Service Manual 7-9
3. Lower the temperature probe into the hose until it contacts the metal screen at
the base of the hose duct adapter. The probe should be approximately centered
on the screen.
4. Test the 34°C setting:
a. Press the 34°C key. Allow the temperature to stabilize for at least 3 minutes.
b. Confirm that the temperature measured by the probe is within 1°C of the
temperature setting.
5. Test the 47°C setting:
a. Press the 47°C key. Allow the temperature to stabilize for at least 3 minutes.
b. Confirm that the temperature measured by the probe is within 1°C of the
temperature setting.
6. Remove the temperature probe from the unit.
7. When finished with all testing:
a. Press the On/Standby key to power off the unit.
b. Unplug the unit from the outlet.
Temperature Accuracy Test Failure
If the test fails, see Basic Troubleshooting on page 6-9.