EasyManua.ls Logo

Icom IC-7100 - D Display

Icom IC-7100
375 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19
MAINTENANCE
19-13
PROBLEM POSSIBLE CAUSE SOLUTION REF.
The antenna is not prop-
erly tuned.
The antenna SWR is adjusted too high.
• Check the coaxial cable.
• Adjust the antenna SWR.
• Change the length of the coaxial cable.
Sec. 6
Sec. 2
The supply voltage de-
creases when transmit-
ting.
The capacity of power supply is too low. Use a power supply with capacity of at
least 22 Amperes.
Sec. 2
The continuous transmis-
sions is inhibited.
The time-out timer activates. Turn OFF the time-out timer function. Sec. 17
Transmitting is impossible
in AM.
The 144/430 MHz frequency band is se-
lected.
• Select the HF/50 MHz frequency band. Sec.3
No voice synthesizer au-
dio from the speaker.
The speech level is set to minimum. Adjust the speech level. Sec. 17
Cloning error occurs when
reading or writing through
the USB cable.
The CI-V data transfer speed is set to 300
or 1200 bps.
The CI-V data transfer speed is set to
Auto, but cloning is performed with 300 or
1200 bps.
Set the CI-V data transfer speed to other
than 300 and 1200 bps.
Select the Clone mode in the Others Set
mode.
Sec. 17
C S -7 1 00
Instruction
manual
Display D
PROBLEM POSSIBLE CAUSE SOLUTION REF.
The displayed frequency
does not change.
The dial lock function is turned ON.
Hold down
SPEECH
for 1 second to turn
OFF the function.
Sec. 5
Scanning D
PROBLEM POSSIBLE CAUSE SOLUTION REF.
A Programmed scan does
not start.
The same frequencies have been pro-
grammed into both “1A–3A” and “1B–3B”
of the scan edge memory channels.
Program different frequencies into the
“1A–3A” and “1B–3B” scan edge memory
channels.
Sec. 12
A Memory scan does not
start.
2 or more memory channels have not
been programmed.
Program more than 2 memory channels. Sec. 11
A Select memory scan
does not start.
2 or more memory channels have not
been designated as select channels.
Designate more than 2 memory channels
as select channels for the scan.
Sec. 12
A Mode select memory
scan does not start.
2 or more memory channels with the de-
sired mode have not been programmed.
Program more than 2 memory channels
with the desired operating mode.
Sec. 11
Troubleshooting
Transmit and receive (Continued) D
Previous view

Table of Contents

Other manuals for Icom IC-7100

Related product manuals