Chapter 5 Parameters Definition
170
• DISC*.CSV (F2 – Test, F8 – Discriminator Plot data)
• PLAT*.CSV (F2 – Test, F9 – Plateau Plot data)
• RESULT*.CSV (Verification Test Result data)
• CALIB*CAL.CSV (Calibration Result data)
• CALIB*CHK.CSV (Calibration Check Result data)
• OPTIMIZE*.CSV (Shift+F2 – Additional Test, F1 – Optimize Gamma HV
data)
• OPTIMIZE*.CSV (Shift+F2 – Additional Test, F2 – Optimize Shelf Shield
Factors data)
Note that both choices are available for F6 – Printout Menu items via the Ctrl+F2
Save As… option (on page 253).
Auto Save Last Contaminated: Choices “Color BMP” and “Color PNG”. Default is
“Color PNG”.
Auto Save All Setup Values: If set to YES (default), a report of all current setup
values is created in an TEXT file named SETUP.TXT in the DATALOG directory
(D:\MonitorProgram\DATALOG\) upon leaving Service mode, overwriting the
previous version of the file, if it exists. A SETUP.TXT file can be created at any
time using the F7 – All Setup Values Printout screen on page 267.
Auto Save Calibration: If set to YES (default), calibration settings are saved in an
TEXT file named CALIB.TXT in the DATALOG directory
(D:\MonitorProgram\DATALOG\) upon leaving the F3 – Calibration screen (on
page 121) if calibration settings were changed, overwriting the previous version of
the file, if it exists. A CALIB.TXT file can be created at any time using the F3 –
Calibration Printout screen on page 257.
Auto Save Calibration Data: If set to YES (default), calibration data are saved in an
TEXT file named CAL_DATA.TXT in the DATALOG directory upon leaving the F3
– Calibration screen if calibration settings were changed, overwriting the previous
version of the file, if it exists. A CAL_DATA.TXT file can be created at any time
using the F4 – Calibration Data Printout screen on page 258.
Auto Save Calibration Check Results: If set to “YES” (default), calibration check
data are saved in an TEXT file named CALCHECK.TXT in the DATALOG
directory. A CALCHECK.TXT files can be created at any time using the F8 –
Calibration Check Results Printout screen on page 268)
Include Derived Sum Zone Efficiencies: Choices are “YES” and “NO” (default).
This affects the CALIB.TXT, CAL_DATA.TXT and CALCHECK.TXT files. If any
of the sum zones have a non-derived efficiency then it will print all the Derived
Efficiencies even if set to “NO”.
Auto Save Calibration Source List: If set to “YES” (default), the calibration source
list is saved in an TEXT file named CAL_LIST.TXT in the DATALOG directory