F3 Setup Menu
Gem-5 User’s Manual 171
upon leaving the Edit Calibration Source List screen (see Edit Calibration Source
List) if the list was changed, overwriting the previous version of the file, if it exists. A
CAL_LIST.TXT file can be created at any time using the F6 Calibration Source
List Printout screen on page 265.
Auto Save Calibration Results: Auto Printout Calibration Results (CSV) files are
produced in the following cases:
Calibration Result D:\MonitorProgram\DATALOG\CALIB
YYYYMMDDHHMMSS CAL.CSV produced automatically (if Auto Save
Calibration Results is set to YES) when leaving the Calibration screen
providing that a Calibration collection was done on at least one detector
while in this screen.
Changing the efficiency to Factory Default or Manual Calibration is also
considered as a Calibration collection for this purpose. Prior to saving,
previous CALIB YYYYMMDDHHMMSS CAL.CSV files are checked
against Keep Calibration Results setting (discussed below) and removed if
appropriate.
Calibration Result D:\MonitorProgram\DATALOG\CALIB
YYYYMMDDHHMMSS CHK.CSV produced automatically (if Auto Save
Calibration Results is set to YES) when leaving Calibration screen providing
that a Calibration Check collection was done on at least one detector while in
this screen. Prior to saving, previous CALIB YYYYMMDDHHMMSS
CHK.CSV files are checked against Keep Calibration Results setting
(discussed below) and removed if appropriate.
The File Printout consists of Title, Column Labels, and Data for each Zone. The data
starts with Channel 1 label in column 1 then a row for each zone If Channel 2 is
present then Channel 2 label is placed in column 1 then a row for each zone
The details of the file contents are as follows:
1
st
row,
“Calibration Results” or “Calibration Check Results”, 2
nd
row blank,
3
rd
row Mirion Technologies (Canberra CA) Ltd. model version, 4
th
row
blank, 5
th
row Monitor ID, Location ID, Serial Number, 6
th
row blank, 7
th
row column labels, 8
th
row column sub labels, units
1
st
column Zone
2
nd
column Description
3
rd
column Type
4
th
column Time Date
5
th
column Efficiency
6
th
column Source Used
7
th
column Emission Rate
8
th
column Activity

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mirion Technologies GEM-5 and is the answer not in the manual?

Mirion Technologies GEM-5 Specifications

General IconGeneral
BrandMirion Technologies
ModelGEM-5
CategoryMonitor
LanguageEnglish

Related product manuals