F3 – Setup Menu
Gem-5 User’s Manual 173
automatically. Each file type is assigned a distinctive filename that includes the date,
time, and whether clean or contaminated, as indicated on the screen.
Defaults are shown in Table 11. A setting of YES allows files to be available if
unexpected results occur.
Table 11 Default Settings for Printout Auto Save items
Auto Save item Default setting
Calibration Results YES
Pre-Extend NO
Detector Check NO
Verification Test YES
Alarm Test YES
For example, for the Verification Test the file name which is auto saved is of form
RESULT YYYYMMDDHHMMSS VT.CSV where YYYY is year, MM is month,
DD is day, HH is hour, MM is minute, SS is second when test sequence was started.
The Keep Verification Test Results settings below determine how long the file is
kept.
In this case, the details of the file contents are as follows:
• 1
st
row title, 2
nd
row blank, 3
rd
row model and software version, 4
th
row blank,
5
th
row Board ID, Location ID, Serial Number, 6
th
row blank, 7
th
row Date
time test sequence started, 8
th
row blank, 9
th
row Result (“Passed” or
“Failed”), 10
th
row blank, 11
th
& 12
th
rows column labels and all subsequent
rows are zones tested
• 1
st
column channel label
• 2
nd
column Zone
• 3
rd
column Zone Description
• 4
th
column Alarm Test Source ID
• 5
th
column Alarm Test Source Activity
• 6
th
column Efficiency
• 7
th
column Expected Count Rate for alarm test source
• 8
th
column Lower Limit Count Rate
• 9
th
column Upper Limit Count Rate
• 10
th
column Background (Average) Count Rate