F2 – Test Menu
Gem-5 User’s Manual 211
The remaining line items are displayed for information only:
Temperature Over 70°C Fault: Indicates status of temperature fault (based on
the readings from the on-board temperature sensor). The fault condition occurs
when the measured temperature is >70°C (or 158°F).
+12V, +5V, and +3.3V: Displays the carrier Board power supply voltages as
indicated by the ADC.
Temperature: Displays the Carrier Board sensor temperature as indicated by the
ADC.
A third column appears in the table beside the
“ID” and “Read” columns called
“Rev”. This should help discern different revisions of the Carrier Board. Contact
the Mirion Technologies (Canberra CA) Ltd. Service affiliate for details.
Function Keys
F3 = Play SINE.WAV: Plays sound file SINE.WAV as a test of the on-board sound
system. Refer to Test Volume setting above.
F4 = Stop SINE.WAV: Stop play of the sound file SINE.WAV.
F6 = Set Factory Defaults: sets the selected item to the factory default value.
Ctrl+F6 = All Factory Defaults: sets all items on this screen to the factory defaults.
Ctrl+F7 = Recall previous Settings: restores the settings in effect when the screen
was opened.
Ctrl+F10 cycles through Normal View, Bench Test View, and Show Hidden View.
Esc = Return to Test Menu