DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument 123
Figure 5‑3 The automatic lift‑off function
5. Press the FREEZE key ( ).
You can now use both of your hands to complete the calibration.
6. Press the ANGLE key ( ). The ANGLE parameter defaults to the coarse
mode. Press the Enter key ( ) in order to switch the ANGLE parameter to the
fine mode. Adjust the ANGLE value with the knob until the lift-off signal is as
close as possible to horizontal (see Figure 5-4 on page 123).
Figure 5‑4 The lift‑off signal as close as possible to horizontal