Olympus nortec 600 - Figure 5-167 the Frequency 2 Scan Signal of the Support Ring; Figure 5-168 the Frequency 2 Scan Signal of the 1.3 MM (0.052 In.) Thru-Wall Hole

Olympus nortec 600
374 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA-10040-01EN, Rev. E, February 2018
Chapter 5240
Figure 5167 The frequency 2 scan signal of the support ring
6. Slowly scan the 1.3 mm (0.052 in.) thru-wall hole only, pressing the ERASE foot
switch as required to clear the screen. When the hole signal is visible on the
NORTEC 600 screen, press the FREEZE key ( ) [see Figure 5-168 on page 240].
Figure 5168 The frequency 2 scan signal of the 1.3 mm (0.052 in.) thruwall hole
7. Press the ANGLE key ( ), and then rotate the frequency 2 signal until the
hole signal reaches a phase of approximately 40°.