Olympus nortec 600 - Figure 5-197 a Slow Scan Rate (Optimal Signal, Left) and a Fast Scan Rate (Loss of Signal, Right)

Olympus nortec 600
374 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA-10040-01EN, Rev. E, February 2018
Chapter 5262
6. Place the probe in a defect-free area of the calibration standard near the thru-wall
flaw, and then press the NULL foot switch.
7. Slowly scan the entire length of the tube, and press the FREEZE key ( ) when
done (see Figure 5-197 on page 262).
Remote-field testing (RFT) requires a slow inspection speed, and practice. Scan the
tube at several inspection speeds and note how the signal varies until you are
comfortable with the scan rate.
Figure 5197 A slow scan rate (optimal signal, left) and a fast scan rate (loss of
signal, right)
8. Adjust the support ring signal so that it is not being clipped by the screen (8 to 9
major divisions) [see Figure 5-198 on page 263].