DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument 259
5.4.3.1 Using the Application
This application uses inspection materials similar to the example illustrated in
Figure 5-135 on page 216 (in the application “Inspecting Heat Exchanger Tubing
Using Dual Frequency — NORTEC 600D Model” on page 215).
The following products are used in this procedure:
• Probe; P/N: TRX-130-300-N20 [U8280123]
• Differential and absolute remote-field probe adaptor; P/N: CBAS-10821-0001
[Q7670054]
• Demonstration calibration standard (P/N: MESX0397 [Q7800046]) and large
support for calibration standard (P/N: MESX0400 [Q7800049]).
• Convenient, highly recommended items (however, not mandatory): foot switch
(P/N: 9522333 [Q7670007]), foot switch armored cable (P/N: 9122404 [Q7670008]),
and foot switch adaptor (P/N: 9522336 [Q2500083])
To set the initial NORTEC 600 configuration
1. Connect the adaptor, probe, and foot-switch adaptor cable to the PROBE
connector on the NORTEC 600.
2. Select CONTINUE (press the A key), and then use the knob to select RFT Pitting
and Wear, and press to accept (see Figure 5-193 on page 259).
Figure 5‑193 The RFT Pitting and Wear application