EasyManua.ls Logo

Olympus nortec 600 - Figure 5-166 the Verified Signals: Thru-Wall Hole (Left) and Support Ring (Right)

Olympus nortec 600
374 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument 239
Figure 5166 The verified signals: thruwall hole (left) and support ring (right)
To calibrate the signals for frequency 2
1. Press the DISP menu key ( ), followed by CHANNEL (B key), and then rotate
the knob until FRQ2 is displayed.
2. Access the MIX menu by pressing the MAIN FILTER menu key ( ) twice to
check the frequency 2 value.
When you use the MIX function, the frequency 2 value must be at least one-half of the
frequency 1 value.
3. Place the probe in a defect-free area near the tube’s support ring, and then press
the NULL foot switch.
4. Scan the support ring and, if necessary, adjust the combined gain (H/V GAIN 2)
of frequency 2 to make sure that the signal does not exceed 80 % of full-screen
height.
5. When you obtain a satisfactory signal, press the FREEZE key ( ) to enable
signal acquisition (see Figure 5-167 on page 240).
NOTE

Table of Contents

Other manuals for Olympus nortec 600

Related product manuals