EasyManua.ls Logo

Olympus nortec 600 - Figure 5-64 the Figure 6 Filter

Olympus nortec 600
374 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA-10040-01EN, Rev. E, February 2018
Chapter 5168
2. Ensure that the scanner motor is turned off before initial instrument nulling, place
the probe in the good hole, and then press the A-LIFT NULL key ( ).
3. Place the probe in the hole with EDM notches, and then adjust the index axis so
that the probe rotates freely next to a defect area.
Leave the scanner in the R position (rotation).
4. Turn on the scanner motor, and then press the MAIN FILTER menu key ( )
twice, and set the SCAN RPM (E key) to 240.
5. Press the MAIN FILTER menu key ( ), and then change the FILT TYP (C key)
to FIG 6 (see Figure 5-64 on page 168).
This procedure uses the Figure 6 filter setting. Olympus recommends using the
Figure 6 filter setting unless an absolute probe is being used, in which case the
Figure 8 filter setting might provide better performance. For more information about
the Figure 6 versus Figure 8 parameters, see “Inspecting Fastener Holes with a
Rotating Scanner — NORTEC 600S and NORTEC 600D Models” on page 126.
Figure 564 The Figure 6 filter
NOTE

Table of Contents

Other manuals for Olympus nortec 600

Related product manuals