DMTA-10040-01EN, Rev. E, February 2018
Chapter 5170
8. Press the GAIN key ( ) twice, and then increase the V GAIN until the notch
signal reaches full screen height (100 % vertical) [see Figure 5-67 on page 170].
Figure 567 Adjusting the vertical gain
9. Press the DISP menu key ( ), and then set the SYNC ANG (D key) to center
the notch signal on the strip chart (see Figure 5-68 on page 170).
Figure 568 Adjusting the sync angle
10. Set the DSP MODE (A key) to WATERFALL.
11. Validate that the PS5 scanner is ready to perform a scan.