6. Press the DISP menu key ( ) again, and then set the POS 2 (C key) to BOT
RGHT.
7. Press the DISP menu key ( ) again, and then set the MIX DISP (A key) to
OFF.
8. Press the ADV SETUP menu key ( ), select ALL SETTINGS (B key), press
the E key, and then the C key. Set the RDG1 CHN to MIX, press the FULL NEXT
key ( ) four times, and then set the RDG2 TYP to OFF. When done, press the
Return key ( ).
The list of dual frequency parameters is shown in Figure 5-116 on page 204.
Figure 5‑116 The list of dual frequency parameters
To calibrate the signals
1. Press the MAIN FILTER menu key ( ), and then set the FREQ (A key) to
9.0 kHz.
2. Press the MAIN FILTER menu key ( ) again, and then set the FREQ 2 (A key)
to 2.2 kHz.
3. Place the probe on position 1 of the standard (see Figure 5-117 on page 205), and
then press the A-LIFT NULL key ( ).
Make sure that the probe remains properly positioned between two fasteners.