Olympus nortec 600 - Figure 5-116 the List of Dual Frequency Parameters

Olympus nortec 600
374 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA-10040-01EN, Rev. E, February 2018
Chapter 5204
6. Press the DISP menu key ( ) again, and then set the POS 2 (C key) to BOT
RGHT.
7. Press the DISP menu key ( ) again, and then set the MIX DISP (A key) to
OFF.
8. Press the ADV SETUP menu key ( ), select ALL SETTINGS (B key), press
the E key, and then the C key. Set the RDG1 CHN to MIX, press the FULL NEXT
key ( ) four times, and then set the RDG2 TYP to OFF. When done, press the
Return key ( ).
The list of dual frequency parameters is shown in Figure 5-116 on page 204.
Figure 5116 The list of dual frequency parameters
To calibrate the signals
1. Press the MAIN FILTER menu key ( ), and then set the FREQ (A key) to
9.0 kHz.
2. Press the MAIN FILTER menu key ( ) again, and then set the FREQ 2 (A key)
to 2.2 kHz.
3. Place the probe on position 1 of the standard (see Figure 5-117 on page 205), and
then press the A-LIFT NULL key ( ).
Make sure that the probe remains properly positioned between two fasteners.