DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument 221
Figure 5‑142 Adjusting the angle of the frequency 2 signal on the support ring
9. Press the GAIN key ( ) four times, and then adjust the H/V GAIN 2 until the
support ring’s frequency 2 signal extends across approximately 10 horizontal
divisions on the screen (see Figure 5-144 on page 222).
Figure 5‑143 Adjusting the gain of the frequency 2 signal on the support ring
10. Press the FREEZE key ( ) to unfreeze the acquisition.
11. Press the MAIN FILTER menu key ( ) repeatedly until the MIX menu is
displayed.