EasyManua.ls Logo

Olympus nortec 600 - Figure 5-183 the ECT Erosion and Corrosion Application

Olympus nortec 600
374 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DMTA-10040-01EN, Rev. E, February 2018
Chapter 5252
2. Select CONTINUE (press the A key), and then use the knob to select ECT Erosion
and Corrosion, and press to accept (see Figure 5-183 on page 252).
Figure 5183 The ECT Erosion and Corrosion application
To calibrate the signals
1. Press FREQ (frequency, A key), and use the knob to select 5kHz.
2. Place the probe in a defect-free area of the calibration standard near the 10 % ID
flaw, and then press the NULL foot switch.
3. Slowly scan the tube until the 10 % ID and 20 % OD flaws are displayed. When
the groove signals are visible on the NORTEC 600 screen, press the FREEZE key
( ) [see Figure 5-184 on page 253].

Table of Contents

Other manuals for Olympus nortec 600

Related product manuals