DMTA-10040-01EN, Rev. E, February 2018
Chapter 5130
Second alternative:
If the surface of your sample is made of the same material as the material of
the hole to be inspected, this alternative is usually more practical and easier.
While the scanner is rotating the probe, simply gently touch the surface of the
sample with the head of the probe, and adjust the signal angle so that the tip
of the signal extends to the right of the impedance plane (see Figure 5-14 on
page 130).
Figure 514 Adjusting the liftoff noise (second alternative)
4. Press the GAIN key ( ), and then decrease the gain with the knob in order to
keep the entire crack signal within the screen area.
A good practice is to set the signal’s maximum deflection to 10 % of the screen’s
horizontal deflection (see Figure 5-15 on page 131).