DMTA-10040-01EN, Rev. E, February 2018
Chapter 5272
Figure 5210 A single exciter probe
Figure 5211 A large defect detected on both sides of a support using the absolute
(ABS) channel
5.4.4.1 Using the Application
This application uses inspection materials similar to the example illustrated in
Figure 5-135 on page 216 (in the application “Inspecting Heat Exchanger Tubing
Using Dual Frequency — NORTEC 600D Model” on page 215).
The following products are used in this procedure:
Probe (P/N: TRS-130-300-N20 [U8280143]
Differential and absolute remote-field probe adaptor; P/N: CBAS-10821-0001
[Q7670054]
Demonstration calibration standard; P/N: MESX0398 [Q7800054]
Convenient, highly recommended items (however, not mandatory): foot switch
(P/N: 9522333 [Q7670007]), foot switch armored cable (P/N: 9122404 [Q7670008]),
and foot switch adaptor (P/N: 9522336 [Q2500083])
Lead pickup (absolute
channel)
Trail pickupExciter