c) Fine-tune the MIX angle (on the MIX page of the MAIN FILTER menu) to
also obtain 40.0 degrees or very close to this value (see Figure 5-147 on
page 224).
Figure 5‑147 Fine‑tuning the MIX angle
2. Press the FULL NEXT key ( ) to activate the full-screen mode, and then scan the
five flat-bottom holes.
The signal should resemble the image shown in Figure 5-148 on page 224.
The list of all parameters is shown in Figure 5-149 on page 225.