6. Press the ANGLE key ( ), and then adjust the lift-off (probe-motion) angle so
that it is as horizontal as possible and the notch signal points upward (see
Figure 5-65 on page 169).
Figure 5‑65 Adjusting the signal angle
7. Press the GAIN key ( ), and then decrease the horizontal gain until the notch
signal is contained within the first horizontal division (10 % horizontal) [see
Figure 5-66 on page 169].
Figure 5‑66 Adjusting the gain