Using the Instrument 157
COATING (C key) is then highlighted on the display (see Figure 5-51 on
page 157).
4. Place the probe on the highest IACS percentage standard with the 4 mil shim, and
then select STORE4 (E key).
Figure 5‑51 The displayed instruction (4)
5. Complete the calibration by selecting DONE after Success is displayed (see
Figure 5-52 on page 157).
Figure 5‑52 The confirmation of a completed calibration