DMTA-10040-01EN, Rev. E, February 2018
Using the Instrument 159
The instrument’s conductivity and coating thickness mode must be activated in order
to access the alarm setups described in this section.
1. Press the ALARM menu key ( ).
2. Press the A key, and then rotate the knob to set the desired COND LO
(conductivity low) value.
3. Press the B key, and then rotate the knob to the desired COND HI (conductivity
high) value.
4. Press the C key, and then set the polarity of the alarm:
Choose POSITIVE to enable an alarm when the conductivity is between the
low and high value.
OR
Choose NEGATIVE to enable an alarm when the conductivity is outside the
low or high value.
5. Press the E key to enable the audible alarm (HORN), and then rotate the knob to
the OFF, LOW, or HIGH setting.
To set the coating thickness alarm
1. Press the ALARM menu key ( ) twice (or, if you have just set up the
conductivity alarm, press only once).
2. Press the A key, and then rotate the knob to the desired THCK LO (low thickness)
value.
3. Press the B key, and then rotate the knob to the desired THCK HI (high thickness)
value.
4. Press the C key, and then set the polarity of the alarm:
Choose POSITIVE to enable an alarm when the coating thickness is between
the low and high value.
OR
Choose NEGATIVE to enable an alarm when the coating thickness is outside
the low or high value.
5. Press the E key to enable the audible alarm (HORN), and then rotate the knob to
the OFF, LOW, or HIGH setting.