Remote control commandsMXO 4 Series
527User Manual 1335.5337.02 ─ 07
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<ProbeCouplingAC> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "AC Coupling" on page 133
PROBe<ch>:SETup:CMOFfset
<CMOffset>
Sets the common-mode offset to compensate for a common DC voltage that is applied
to both input sockets (referenced to the ground socket). The setting is available for
Rohde & Schwarz differential probes and for modular probes in CM measurement
mode.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<CMOffset> Range: -1E+26 to 1E+26
Increment: 0.001
*RST: 0
Default unit: V
Usage: Asynchronous command
Manual operation: See "CM offset" on page 132
PROBe<ch>:SETup:DISPlaydiff <DisplayDiff>
Selects the voltage to be measured by the R&S ProbeMeter of differential active
probes:
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<DisplayDiff> DIFFerential | SINGleended
DIFFerential
Measures differential and common mode voltages
SINGleended
Measures the voltage between the positive/negative signal
socket and the ground.
*RST: DIFFerential
Usage: Asynchronous command
Manual operation: See "Display" on page 132
Acquisition and setup