Remote control commandsMXO 4 Series
535User Manual 1335.5337.02 ─ 07
17.8.8.4 Settings for current probes
The probe suffix <ch> selects the input channel to which the probe is connected.
PROBe<ch>:SETup:GAIN:AUTO?...................................................................................535
PROBe<ch>:SETup:GAIN:MANual.................................................................................. 535
PROBe<ch>:SETup:DEGauss.........................................................................................535
PROBe<ch>:SETup:OFFSet:ZADJust..............................................................................536
PROBe<ch>:SETup:OFFSet:STPRobe............................................................................ 536
PROBe<ch>:SETup:GAIN:AUTO?
Returns the gain of a detected or predefined current probe.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Return values:
<Gain> Range: 0.001 to 1000
Increment: 0.1
*RST: 1
Default unit: A/V
Usage: Query only
Asynchronous command
Manual operation: See "Gain, Manual gain" on page 136
PROBe<ch>:SETup:GAIN:MANual <Gain>
Sets the gain for an unknown current probe.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<Gain> Range: 0.0001 to 10000
Increment: 0.0001
*RST: 1
Default unit: V/V
Usage: Asynchronous command
Manual operation: See "Gain, Manual gain" on page 136
PROBe<ch>:SETup:DEGauss
Demagnetizes the core if it has been magnetized by switching the power on and off, or
by an excessive input. Always carry out demagnetizing before measurement.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Acquisition and setup