Remote control commandsMXO 4 Series
536User Manual 1335.5337.02 ─ 07
Usage: Event
Asynchronous command
Manual operation: See "DeGauss" on page 137
PROBe<ch>:SETup:OFFSet:ZADJust <ZeroAdjustValue>
Set the waveform to zero position. It corrects the effect of a voltage offset or tempera-
ture drift. To set the value by the instrument, use PROBe<ch>:SETup:OFFSet:AZERo
.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Parameters:
<ZroAdjVal> Range: -100 to 100
Increment: 0.1
*RST: 0
Default unit: %
Usage: Asynchronous command
Manual operation: See "Zero adjust" on page 137
PROBe<ch>:SETup:OFFSet:STPRobe
Saves the zero adjust value in the probe box. If you connect the probe to another
channel or to another Rohde & Schwarz oscilloscope, the value is read out again, and
you can use the probe without further adjustment.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Usage: Event
Asynchronous command
Manual operation: See "Save to probe" on page 137
17.8.8.5 Probe attributes
PROBe<ch>:ID:PARTnumber?........................................................................................536
PROBe<ch>:ID:PRDate?................................................................................................537
PROBe<ch>:ID:SRNumber?........................................................................................... 537
PROBe<ch>:ID:SWVersion?........................................................................................... 537
PROBe<ch>:SETup:CAPacitance?.................................................................................. 537
PROBe<ch>:SETup:IMPedance?.................................................................................... 538
PROBe<ch>:ID:PARTnumber?
Queries the Rohde & Schwarz part number of the probe.
Suffix:
<ch>
.
1 to 4, index of the analog channel
Acquisition and setup