Remote control commandsMXO 4 Series
577User Manual 1335.5337.02 ─ 07
Manual operation: See "Position range" on page 179
LAYout<ly>:ZOOM<zo>:HORizontal:RELative:SPAN <RelativeSpan>
Defines the width of the zoom area in relative values.
Suffix:
<ly>
.
1...4, index of the SmartGrid layout
<zo> 1...4, index of the zoom
Parameters:
<RelativeSpan> Range: 1E-15 to 100
Increment: 0.1
*RST: 1
Default unit: %
Usage: Asynchronous command
Manual operation: See "Position range" on page 179
LAYout<ly>:ZOOM<zo>:HORizontal:RELative:STARt <RelativeStart>
Defines the lower limit of the zoom area on the x-axis in relative values.
Suffix:
<ly>
.
1...4, index of the SmartGrid layout
<zo> 1...4, index of the zoom
Parameters:
<RelativeStart> Range: 0 to 100
Increment: 0.1
*RST: 0
Default unit: %
Example: See Chapter 17.4.1, "SmartGrid layout with zoom", on page 474.
Usage: Asynchronous command
Manual operation: See "Start Stop" on page 179
LAYout<ly>:ZOOM<zo>:HORizontal:RELative:STOP <RelativeStop>
Defines the upper limit of the zoom area on the x-axis in relative values.