EasyManua.ls Logo

Rohde & Schwarz MXO 4 Series - Page 584

Rohde & Schwarz MXO 4 Series
925 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commandsMXO 4 Series
584User Manual 1335.5337.02 ─ 07
Example:
CALCulate:MATH3:LABel 'M3AddC1C2'
Usage: Asynchronous command
Manual operation: See "Label" on page 184
CALCulate:MATH<m>:VERTical:SCALe[:VALue] <VerticalScale>
Sets the scale of the y-axis in the math function diagram. The value is defined as
"<unit> per division", e.g. 50
mV/div. In this case, the horizontal grid lines are displayed
in intervals of 50 mV.
Suffix:
<m>
.
1...5, index of the math waveform
Parameters:
<VerticalScale> Range: 1E-12 to 100E+12
Increment: 1E-05
*RST: 0.5
Default unit: V
Usage: Asynchronous command
Manual operation: See "[Scale]" on page 43
CALCulate:MATH<m>:UNIT <UserUnit>
Sets a user-defined unit for the math operation.
Suffix:
<m>
.
1...5, index of the math waveform
Parameters:
<UserUnit> String with the base unit
Example:
CALCulate:MATH:UNIT 'A'
Usage: Asynchronous command
Manual operation: See "Unit" on page 191
CALCulate:MATH<m>:VERTical:OFFSet <VerticalOffset>
Sets a voltage offset to adjust the vertical position of the math function on the screen.
Negative values move the waveform up, positive values move it down.
Suffix:
<m>
.
1...5, index of the math waveform
Parameters:
<VerticalOffset> Range: -100E12 to 100E12
Increment: 0.01
*RST: 0
Default unit: div
Waveform analysis

Table of Contents