Remote control commandsMXO 4 Series
638User Manual 1335.5337.02 ─ 07
Parameters:
<YPosition2> Range: -50 to 50
Increment: 0.01
*RST: 0
Usage: Asynchronous command
Manual operation: See "Y1 position, Y2 position" on page 213
CURSor<cu>:SSCReen
Resets the cursors to their initial positions. Reset is helpful if the cursors have disap-
peared from the display or need to be moved for a larger distance.
Suffix:
<cu>
.
1...2, index of the cursor set
Usage: Setting only
Asynchronous command
Manual operation: See "Place on display" on page 213
CURSor<cu>:TRACking[:STATe] <TrackCurve>
If enabled, the horizontal cursor lines tracks the waveform.
Suffix:
<cu>
.
1...2, index of the cursor set
Parameters:
<TrackCurve> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "Track waveform" on page 213
CURSor<cu>:LABel <ShowLabel>
Shows the cursor labels in the diagram.
Suffix:
<cu>
.
1...2, index of the cursor set
Parameters:
<ShowLabel> ON | OFF
*RST: OFF
Usage: Asynchronous command
Manual operation: See "Show label" on page 214