Remote control commandsMXO 4 Series
686User Manual 1335.5337.02 ─ 07
Manual operation: See "Current" on page 282
POWer<m>:HARMonics:SOURce[:VOLTage] <VoltageSource>
Sets the channel for the voltage source input of the power harmonics analysis.
Suffix:
<m>
.
1...3, index of the power analysis measurement
Parameters:
<VoltageSource> C1 | C2 | C3 | C4
Usage: Asynchronous command
Manual operation: See "Voltage" on page 282
POWer<m>:HARMonics:STANdard
<Standard>
Sets a standard for the current harmonic measurement.
Suffix:
<m>
.
1...3, index of the power analysis measurement
Parameters:
<Standard> ENA | ENB | ENC | END | MIL | RTCA
ENA: EN 61000-3-2 Class A
ENB: EN 61000-3-2 Class B
ENC: EN 61000-3-2 Class C
END: EN 61000-3-2 Class D
MIL: MIL-STD-1399
RTCA: RTCA DO-160
*RST: ENA
Usage: Asynchronous command
Manual operation: See "Standard" on page 282
POWer<m>:HARMonics:FREQuency:EN <FundFreqEN61000>
Sets the fundamental frequency of the input signal for the EN61000 standard, if
POWer<m>:HARMonics:STANdard is set to ENA / ENB / ENC / END.
Suffix:
<m>
.
1...3, index of the power analysis measurement
Parameters:
<FundFreqEN61000> F50 | F60 | AUTO
F50: 50 Hz
F60: 60 Hz
AUTO: automatically set
*RST: AUTO