Remote control commandsMXO 4 Series
687User Manual 1335.5337.02 ─ 07
Usage: Asynchronous command
Manual operation: See "Fund. Freq" on page 282
POWer<m>:HARMonics:FREQuency:MIL <FundFreqMIL>
Sets the fundamental frequency of the input signal for the MIL standard, if POWer<m>:
HARMonics:STANdard is set to MIL.
Suffix:
<m>
.
1...3, index of the power analysis measurement
Parameters:
<FundFreqMIL> F60 | F400
F60: 60 Hz
F400: 400 Hz
*RST: F60
Usage: Asynchronous command
Manual operation: See "Fund. Freq" on page 282
POWer<m>:HARMonics:FREQuency:RTCA <FundamFreqRTCA>
Sets the fundamental frequency of the input signal for the RTCA standard, if
POWer<m>:HARMonics:STANdard is set to RTCA.
Suffix:
<m>
.
1...3, index of the power analysis measurement
Parameters:
<FundFreqRTCA> F400 | NVF | WVF
F400: 400 Hz
NVF
WVF
*RST: F400
Usage: Asynchronous command
Manual operation: See "Fund. Freq" on page 282
POWer<m>:HARMonics:STATistics:RESet
Resets the statistics for the power harmonics analysis.
Suffix:
<m>
.
1...3, index of the power analysis measurement
Usage: Setting only
Asynchronous command
Manual operation: See "Clear Results" on page 283