Remote control commandsMXO 4 Series
689User Manual 1335.5337.02 ─ 07
Suffix:
<m>
.
1...3, index of the power analysis measurement
Parameters:
<Scaling> LIN | LOG
*RST: LIN
Usage: Asynchronous command
Manual operation: See "Scale bar graph" on page 282
Power harmonics results
POWer<m>:HARMonics:AVAilable?
Returns the number of measured harmonics.
Suffix:
<m>
.
1...3, index of the power analysis measurement
Return values:
<Count>
Usage: Query only
Asynchronous command
POWer<m>:HARMonics:MEASurement:THDFundament[:ACTual]?
Returns the total harmonic distortion relative to fundamental (THDF).
Suffix:
<m>
.
1...3, index of the power analysis measurement
Usage: Query only
Asynchronous command
POWer<m>:HARMonics:MEASurement:THDRms[:ACTual]?
Returns the total harmonic distortion relative to the RMS amplitude (THD RMS).
Suffix:
<m>
.
1...3, index of the power analysis measurement
Usage: Query only
Asynchronous command
POWer<m>:HARMonics:RESult<n>:VIOLation?
Queries if the value of n-th harmonic is within the defined limit for the current measure-
ment.
Suffix:
<m>
.
1...3, index of the power analysis measurement
Applications