Remote control commandsMXO 4 Series
748User Manual 1335.5337.02 ─ 07
Manual operation: See "Packets" on page 357
SBUS<sb>:UART:PACKets <FrmSeparation>
Defines the method of packet separation. A packet is a number of subsequent words in
a date stream.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<FrmSeparation> NONE | TOUT
NONE
Packets are not considered.
EWORd
End word, the end condition of a packet is a pattern. To define
the end word, use SBUS<m>:UART:EWORd.
TOUT
Defines a timeout between the packets. To set the timeout, use
SBUS<sb>:UART:TOUT.
*RST: TOUT
Usage: Asynchronous command
Manual operation: See "Packets" on page 357
SBUS<sb>:UART:PARity <Parity>
Defines the optional parity bit that is used for error detection.
Suffix:
<sb>
.
1...4, index of the serial bus
Parameters:
<Parity> NONE | ODD | EVEN | MARK | SPC | DC
MARK
The parity bit is always a logic 1.
SPC
SPaCe: The parity bit is always a logic 0.
DC
Do not care: the parity is ignored.
*RST: NONE
Usage: Asynchronous command
Manual operation: See "Parity" on page 357