EasyManua.ls Logo

Rohde & Schwarz MXO 4 Series - Page 757

Rohde & Schwarz MXO 4 Series
925 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote control commandsMXO 4 Series
757User Manual 1335.5337.02 ─ 07
SBUS<sb>:UART:FILTer:ERENable <ErrorName>,<Enabler>
SBUS<sb>:UART:FILTer:ERRor<n>:ENABle <Enable>
Defines the error type to be filtered on.
Suffix:
<sb>
.
1...4, index of the serial bus
<n> index of the error
Parameters:
<Enable> ON | OFF
*RST: ON
Usage: Asynchronous command
Manual operation: See "Error type" on page 361
SBUS<sb>:UART:FILTer:DOPerator <Frame>,<Field>,<Operator>
SBUS<sb>:UART:FILTer:FRAMe<fr>:FLD<fl>:DOPerator <Data_Operator>
Sets the operator for the data pattern in the selected field of the selected frame.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
<fl> index of the field
Parameters:
<Data_Operator> EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
*RST: EQUal
Usage: Asynchronous command
Manual operation: See "Edit" on page 360
SBUS<sb>:UART:FILTer:DMIN <Frame>,<Field>,<Data>
SBUS<sb>:UART:FILTer:FRAMe<fr>:FLD<fl>:DMIN <Data_Min>
Specifies the data pattern, or sets the start value of a data pattern range.
Suffix:
<sb>
.
1..4
Selects the serial bus.
<fr> *
Specifies the frame number.
<fl> *
Specifies the field number within the frame.
Parameters:
<Data_Min>
Protocols

Table of Contents