Remote control commandsMXO 4 Series
756User Manual 1335.5337.02 ─ 07
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
<fl> index of the field
Parameters:
<Index_Min> Range: 1 to 65535
Increment: 1
*RST: 1
Usage: Asynchronous command
Manual operation: See "Edit" on page 360
SBUS<sb>:UART:FILTer:IMAX
<Frame>,<Field>,<Data>
SBUS<sb>:UART:FILTer:FRAMe<fr>:FLD<fl>:IMAX <Index_Max>
Sets the end value of an index range if the operator is set to INRange.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
<fl> index of the field
Parameters:
<Index_Max> Range: 1 to 65535
Increment: 1
*RST: 65535
Usage: Asynchronous command
Manual operation: See "Edit" on page 360
SBUS<sb>:UART:FILTer:IOPerator <Frame>,<Field>,<Operator>
SBUS<sb>:UART:FILTer:FRAMe<fr>:FLD<fl>:IOPerator <Index_Operator>
Sets the operator for the index in the selected field of the selected frame.
Suffix:
<sb>
.
1...4, index of the serial bus
<fr> index of the frame
<fl> index of the field
Parameters:
<Index_Operator> EQUal | INRange | RANGe
*RST: INRange
Usage: Asynchronous command
Manual operation: See "Edit" on page 360