Remote control commands
R&S
®
RTE
1118User Manual 1326.1032.02 ─ 20
CHANnel<m>:SCALe <Scale>
Sets the vertical scale for the indicated channel.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Scale> Scale value, given in Volts per division.
Range: Depends on attenuation factors, coupling, and
instrument model, see below.
Increment: Depends on vertical and probe settings
*RST: 0.05
Default unit: V/div
With 1:1 probe, external attenuation = 1, and 50 Ω input cou-
pling, the vertical scale (input sensitivity) is minimum 1 mV/div
(RTE 1317.2500) or 0.5 mV/div (RTE 1326.2000) to maximum 1
V/div. For 1 MΩ input coupling, the upper limit is 10 V/div. If the
probe and/or external attenuation is changed, multiply the values
by the attenuation factors to get the actual scale range.
Usage:
Asynchronous command
CHANnel<m>:RANGe <Range>
Sets the voltage range across the 10 vertical divisions of the diagram. Use the com-
mand alternativly instead of CHANnel<m>:SCALe.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Range> Voltage range value
Range: Dependencies see below.
Increment: Depends on vertical and probe settings
*RST: 0.5
Default unit: V/div
Depends on attenuation factors and coupling. With 1:1 probe,
external attenuation = 1 and 50 Ω input coupling, the range is 10
mV to 10 V. For 1 MΩ input coupling, it is 10 mV to 100 V. If the
probe and/or external attenuation is changed, multiply the range
values by the attenuation factors.
Usage: Asynchronous command
CHANnel<m>:POSition <Position>
Sets the vertical position of the indicated channel as a graphical value.
Acquisition and setup