Remote control commands
R&S
®
RTE
1259User Manual 1326.1032.02 ─ 20
<MeasType>
parameter value
(remote control)
Measurement
(manual opera-
tion)
Description, result Suffix
<n>
DTOTrigger Delay to trigger Time between the trigger event and a following signal
slope. High accuracy even if the trigger event is outside
the acquisition data.
46
PROBemeter Trig. ProbeMeter DC voltage measured by the connected active R&S
probe
47
SLERising Slew rate on ris-
ing edge
Steepness of the rising edge: voltage difference between
the lower and higher reference level, divided by the rise
time.
48
SLEFalling Slew rate on fall-
ing edge
Steepness of the falling edge: voltage difference
between the higher and lower reference level, divided by
the fall time.
49
not used 50
MEASurement<m>:ENVSelect...................................................................................... 1259
MEASurement<m>:DETThreshold.................................................................................1260
MEASurement<m>:AMPTime:ALEVel............................................................................ 1260
MEASurement<m>:AMPTime:PFSLope......................................................................... 1260
MEASurement<m>:AMPTime:PSLope........................................................................... 1261
MEASurement<m>:AMPTime:DELay<n>:DIRection........................................................ 1261
MEASurement<m>:AMPTime:DELay<n>:ECOunt........................................................... 1262
MEASurement<m>:AMPTime:DELay<n>:SLOPe............................................................ 1262
MEASurement<m>:AMPTime:PTCount.......................................................................... 1263
MEASurement<m>:AMPTime:ESLope........................................................................... 1263
MEASurement<m>:AMPTime:CSLope........................................................................... 1263
MEASurement<m>:AMPTime:CLCK<n>:LSELect........................................................... 1264
MEASurement<m>:AMPTime:DATA<n>:LSELect............................................................1264
MEASurement<m>:AMPTime:DTOTrigger<n>:SLOPe.....................................................1264
MEASurement<m>:AMPTime:DTOTrigger<n>:LSELect................................................... 1265
MEASurement<m>:AMPTime:LCHeck<n>:VALid............................................................ 1265
MEASurement<m>:AMPTime:LCHeck<n>:LOWer:LIMit...................................................1265
MEASurement<m>:AMPTime:LCHeck<n>:UPPer:LIMit................................................... 1265
MEASurement<m>:AMPTime:LCHeck<n>:LOWer:MARGin............................................. 1266
MEASurement<m>:AMPTime:LCHeck<n>:UPPer:MARGin.............................................. 1266
MEASurement<m>:ENVSelect <EnvelopeCurve>
The command is only relevant for measurements on envelope waveforms. It selects
the envelope to be used for measurement.
Suffix:
<m>
.
1..10
See "Selection of the measurement: MEASurement<m>"
on page 1249.
Automatic measurements