Remote control commands
R&S
®
RTE
1325User Manual 1326.1032.02 ─ 20
Usage: Asynchronous command
CALCulate:MATH<m>:FFT:MAGNitude:RANGe <Range>
Defines the vertical value range in spectrum mode.
Suffix:
<m>
.
1..8
math waveform
Parameters:
<Range> Range: 1 to 500
Increment: 1
*RST: 100
Default unit: dB
Usage: Asynchronous command
CALCulate:MATH<m>:FFT:MAGNitude:SCALe <MagnitudeScale>
Defines the scaling of the y-axis. The display values are valid for 50Ω termination
impendance.
For details on the available scaling modes, see "Magnitude unit" on page 414.
Suffix:
<m>
.
1..8
math waveform
Parameters:
<MagnitudeScale> LINear | DBM | DB | DBUV | DBMV | DBV | DBPS | DBNS |
DBUS | DBMS | DBS | DBHZ | DBKHZ | DBKHz | DBMHZ |
DBMHz | DBGHZ | DBGHz | DBA | DBMA | DBUA
LINear
Linear scaling; displays the RMS value of the voltage
*RST: DBM
Usage: Asynchronous command
Table 18-10: Logarithmic scaling values
DBM dBm
DB dB (related to reference level)
DBUV dBμV
DBMV dBmV
DBV dBV
DBPS dBps
DBNS dBns
DBUS dBμs
Spectrum analysis