Remote control commands
R&S
®
RTE
1534User Manual 1326.1032.02 ─ 20
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one data pattern
to be set with TRIGger<m>:LIN:DMIN.
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with TRIGger<m>:LIN:DMIN and TRIGger<m>:
LIN:DMAX
*RST: EQUal
TRIGger<m>:LIN:DMIN
<DataPattern>
Specifies a data pattern, or sets the the start value of a data pattern range.
Parameters:
<DataPattern> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:LIN:DMAX <DataPatternTo>
Sets the the end value of an data range if TRIGger<m>:LIN:DCONdition is set to
INRange or OORange.
Parameters:
<DataPatternTo> Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076. The string parameter accepts the bit
value X (don't care).
TRIGger<m>:LIN:BORDer <Endianness>
Sets the byte order (endianness) of the data transfer.
According to the standard, LIN data is transmitted in little endian transfer order.
Parameters:
<Endianness> BENDian | LENDian
BENDian
Big endian, data is analyzed and evaluated in the order of
reception.
LENDian
Little endian, the instrument reads the complete data, reverses
the byte order of the data, and compares it with the specified
data word.
*RST: BENDian