Remote control commands
R&S
®
RTE
1566User Manual 1326.1032.02 ─ 20
TRIGger<m>:FLXRay:FMIN <IdPattern>
Specifies a frame identifier pattern - the number of the slot - or sets the the start value
of an identifier range.
Parameters:
IdPattern Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076.
TRIGger<m>:FLXRay:FMAX <IdPatternTo>
Sets the the end value of an identifier range if the condition TRIGger<m>:FLXRay:
FCONdition is set to INRange or OORange.
Parameters:
IdPatternTo Numeric or string pattern, see Chapter 18.4.6, "Bit pattern
parameter", on page 1076.
FMAX must be greater or equal than FMIN, and the position of
the X bits are common to FMIN and FMAX.
TRIGger<m>:FLXRay:PCONdition <PayloadLengthOp>
Sets the operator for the payload length trigger setting. You can defined an exact
value, or a range.
Parameters:
<PayloadLengthOp> OFF | ANY | EQUal | NEQual | LTHan | LETHan | GTHan |
GETHan | INRange | OORange
OFF = ANY
The payload length is not relevant for the trigger condition.
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one payload
length to be set with TRIGger<m>:FLXRay:PMIN.
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with TRIGger<m>:FLXRay:PMIN and TRIGger<m>:
FLXRay:PMAX.
*RST: 
OFF
TRIGger<m>:FLXRay:PMIN <PayloadLength>
Specifies a payload length - the number of words in the payload segment - or sets the
the start value of an payload length range. Information is transmitted in 2-byte words,
so the number of data bytes in the payload segment is twice the payload length.
Protocols