Remote control commands
R&S
®
RTE
1567User Manual 1326.1032.02 ─ 20
Parameters:
<PayloadLength> Range:  0  to  127
Increment:  1
*RST:  0
TRIGger<m>:FLXRay:PMAX <PayloadLengthTo>
Sets the the end value of a payload length range if the condition TRIGger<m>:
FLXRay:PCONdition is set to INRange or OORange.
Parameters:
<PayloadLengthTo> Range:  0  to  127
Increment:  1
*RST:  0
TRIGger<m>:FLXRay:CENable <CycleCount>
Sets the operator to define a cycle count or a cycle count range.
Parameters:
<CycleCount> OFF | EQUal | NEQual | LTHan | LETHan | GTHan | GETHan |
INRange | OORange
OFF
The cycle count is not relevant for the trigger condition.
EQUal | NEQual | LTHan | LETHan | GTHan | GETHan
Equal, Not equal, Less than, Less or equal than, Greater Than,
Greater or equal than. These conditions require one cycle count
number to be set with TRIGger<m>:FLXRay:CMIN.
INRange | OORange
In range / Out of range: Set the minimum and maximum value of
the range with TRIGger<m>:FLXRay:CMIN and TRIGger<m>:
FLXRay:CMAX.
*RST: 
OFF
TRIGger<m>:FLXRay:CMIN <CycleCountMin>
Specifies a cycle count - the number of the current FlexRay cycle - or sets the the start
value of an cycle count range.
Parameters:
<CycleCountMin> Range:  0  to  63
Increment:  1
*RST:  0
TRIGger<m>:FLXRay:CMAX <CycleCountMax>
Sets the the end value of a cycle count range if the condition TRIGger<m>:FLXRay:
CENable on page 1567 is set to INRange or OORange.
Protocols